首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Metal On‐Line Voltammetric Analyzer for On‐Site Monitoring of Iron Speciation in Acid Mine Drainage Waters: Development and Characterization
Authors:T Hezard  G Durand  M Rakib  Ph Viers  S Roy  N Guigues  M Brach  J‐P Heitzmann  P Eberlé  F Bellouard
Institution:1. Laboratoire de Génie des Procédés et Matériaux, Ecole Centrale Paris, Grande Voie des Vignes, 92295?Chatenay‐Malabry, France;2. BRGM, Service Métrologie, Monitoring, Analyse, 3 Avenue Claude Guillemin, BP 6009, 45060 Orléans Cedex 02, France;3. Heito SME, 5 rue Gramme, 75015?Paris, France
Abstract:The development of a simple, low cost, portable metal on‐line voltammetric analyzer (MOVA) for the iron analysis, as well as other species, in mining effluents is presented. It consists of a voltammetric cell based on an impinging jet flow configuration, a fluidic system controlled by gravity and electronics. Laboratory tests were performed to optimize the system as well as the analytical conditions for Fe speciation measurements. MOVA was then tested in the laboratory, in samples of both mining effluents. The results showed that iron in both mining effluents, as well as copper and arsenic, can be measured in oxygenated samples.
Keywords:Iron  Mining effluents  On‐site device  Impinging jet electrode  Voltammetry
设为首页 | 免责声明 | 关于勤云 | 加入收藏