方波激发非接触式电导检测器信噪比的影响因素研究
引用本文:张水锋,王立世,党志,黄新建,刘笑笑,肖明威.方波激发非接触式电导检测器信噪比的影响因素研究[J].分析试验室,2008,27(2):10-14.
作者姓名:张水锋  王立世  党志  黄新建  刘笑笑  肖明威
作者单位:华南理工大学化学科学学院,广州,510640
基金项目:国家自然基金(20475018),广东省自然科学基金(07006544)项目资助
摘    要:研制了用于毛细管电泳的方波激发非接触式电导检测器(CCD)并详细考查了对检测器信噪比影响较大的几个重要因素,包括激发频率和运算放大器、基线分压电路和A/D转换器、数字低通滤波;以及常见缓冲体系2-(N-吗啡啉)-乙磺酸/组氨酸(MES/His)体系中电解质浓度,添加剂十六烷基三甲基溴化铵(CTAB)对金属离子的检出信号的影响。所得的结果对于该类检测器的设计和优化具有参考价值。

关 键 词:非接触式电导检测器  方波  毛细管电泳  信噪比  金属离子
文章编号:1000-0720(2008)02-010-05
收稿时间:2006-12-20
修稿时间:2007-02-26

Factors effecting on signal-to-noise ratio of square wave excitation contactless conductivity detection
ZHANG Shui-feng,WANG Li-shi,DANG Zhi,HUANG Xin-jian,LIU Xiao-xiao,XIAO Ming-wei.Factors effecting on signal-to-noise ratio of square wave excitation contactless conductivity detection[J].Chinese Journal of Analysis Laboratory,2008,27(2):10-14.
Authors:ZHANG Shui-feng  WANG Li-shi  DANG Zhi  HUANG Xin-jian  LIU Xiao-xiao  XIAO Ming-wei
Abstract:A novel square wave excitation contactless conductivity detection was presented, and some significant factors effecting on the signal-to-noise(S/N) ratio of the detector were examined in detail.The assessments of S/N ratios were carried out,including the optimization of the excitation frequency of the square wave and the performance of the operational amplifier,both the parameters of the background baseline voltage-limited resistance and the A/D converter,the digital low-pass filtering parameters,as well as the concentration of the common running buffer,2-(N-morpholino) ethanesulfonic acid/L-Hisdine(MES/His) and the electroosmosis flow modified reagent,e.g.cetyltrimethylammonium bromide(CTMAB).The obtained results are significant references for optimization of the contactless conductivity detector.
Keywords:Cootactless conductivity detector  Square wave  Capillary electrophoresis  Signal-to-noise ratio  Metal ions