On-line Preconcentration and Determination of Trace Metals in Water Samples by Flow Injection Combined with Flame Atomic Absorption Spectrometry via Calix[4]arene Carboxylic Acid Packed Micro-column
Authors:GUO Da-wang  ZHAO Xiao-wei  JIA Qiong  ZHOU Wei-hong
Institution:College of Chemistry, Jilin University, Changchun 130012, P. R. China
Abstract:An adsorbent calix4]arene carboxylic acid was employed as the adsorption material for on-line flow injection( FI) micro-column preconcentration coupled with flame atomic absorption spectrometry(FAAS) determination of trace heavy metals(Cu, Pb, Co, Ni and Cd). Parameters such as the pH, loading time and flow rate of sample, and the concentration, volume and flow rate of eluent were optimized. The enrichment factors are 50.0, 56.5, 11.6, 12.1 and 19.1 for Cu, Pb, Co, Ni, and Cd, respectively, and a sample throughput of 20 h-1 was obtained. The limits of detection for Cu, Pb, Co, Ni, and Cd were in a range of 1.56―3.91 μg/L, and the relative standard deviations(RSDs) were less than 2.76%(n=7). Furthermore, the proposed method was successfully applied to the determination of Cu, Pb, Co, Ni, and Cd in certified reference materials and various water samples.
Keywords:Calix[4]arene carboxylic acid  Preconcentration  Flow injection  Metal ion  Flame atomic absorption spectrometry(FAAS)  