Investigation of commercial sorbents for the analysis of opioid peptides in human plasma by on‐line SPE‐CE
Authors:Fernando Benavente  Silvia Medina‐Casanellas  José Barbosa  Victoria Sanz‐Nebot
Institution:Department of Analytical Chemistry, University of Barcelona, Barcelona, Spain
Abstract:In this study, we investigated the performance of several commercial sorbents (Sep‐pack® C18, tC18, C8 and tC2, Oasis® HLB, Isolute® ENV+, Strata?‐X and Oasis® MCX) for the determination of opioid peptides by solid‐phase extraction coupled on‐line to capillary electrophoresis (SPE‐CE). First, standard solutions were analyzed in order to achieve the lowest LOD and the best electrophoretic separations using UV detection. The best results were obtained using C18, C8 and tC2 sorbents, which were examined for the analysis of spiked human plasma samples. A double‐step sample clean‐up pretreatment, which consisted of precipitation with acetonitrile and filtration, was needed to prevent saturation of the on‐line SPE microcartridge. The filtration step was critical to obtain optimum analyte recovery and to clean up the sample matrix. A range of centrifugal filters and filtration conditions were tested and the recoveries of the sample pretreatment were evaluated by CE‐ESI‐MS. The LODs attained through SPE‐CE‐UV were approximately ten‐fold better with C18 than with C8 and tC2. The 0.1 μg/mL LODs achieved by C18‐SPE‐CE‐UV were further improved until we could detect 1 ng/mL concentrations of opioid peptides in plasma samples by C18‐SPE‐CE‐ESI‐MS, due to the outstanding selectivity of the MS detection.
Keywords:MS  Neuropeptides  On‐line  Preconcentration  Plasma  UV