Determination of Copper and Aluminium in Steel by Means of Hollow Cathode and Glow Discharge Light Sources: A Comparative Study
Authors:S Caroli  A Alimonti  P Delle Femmine
Institution:Laboratorio di Tossicologia , Istituto Superiore di Sanità. Viale Regina Elena , 299. 00161, Rome, Italy
Abstract:Abstract

The applicability of hollow cathode and glow discharge light sources to the determination of minor constituents in steel was tested in the case of copper (0.033 ? 0.34wt]%) and aluminium (0.015 ? 0. 16%). The linearity of the calibration graphs, reflected by the Bravais-Pearson coefficient, was found to be slightly improved in all the cases investigated when operating with hollow cathode. Furthermore, results showed that blackening values of the analytical lines can be plotted directly without resorting to reference lines.
Keywords:Hollow Cathode  Glow Discharge  Emission Spectroscopy  Steel  Aluminium  Copper