首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of on-line preconcentration and flow-injection amperometry for phosphate determination in fresh and marine waters
Authors:Udnan Y  McKelvie I D  Grace M R  Jakmunee J  Grudpan K
Institution:a Department of Chemistry, Faculty of Science, Chiang Mai University, Chiang Mai 50200, Thailand
b Water Studies Centre, School of Chemistry, Monash University, P.O. Box 23, Vic. 3800, Australia
c Institute for Science and Technology Research and Development, Chiang Mai University, Chiang Mai 50200, Thailand
Abstract:Dissolved reactive phosphorus (DRP) was determined as orthophosphate (PO4-P) in fresh and saline water samples by flow-injection (FI) amperometry, without and with in-valve column preconcentration. Detection is based on reduction of the product formed from the reaction of DRP with acidic molybdate at a glassy carbon working electrode (GCE) at 220 mV versus the Ag/AgCl reference electrode. A 0.1 M potassium chloride solution was used as both supporting electrolyte and eluent in the preconcentration system. For the FI configuration without preconcentration, a detection limit of 3.4 μg P l−1 and sample throughput of 70 samples h−1 were achieved. The relative standard deviations for 50 and 500 μg P l−1 orthophosphate standards were 5.2 and 5.9%, respectively. By incorporating an ion exchange preconcentration column, a detection limit of 0.18 μg P l−1 was obtained for a 2-min preconcentration time (R.S.D.s for 0.1 and 1 μg P l−1 standards were 22 and 1.0%, respectively). Potential interference from silicate, sulfide, organic phosphates and sodium chloride were investigated. Both the systems were applied to the analysis of certified reference materials and water samples.
Keywords:Phosphate  Flow-injection amperometry  Preconcentration  Bio-Rad AG-X8  DRP  Schlieren effect
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号