ABTS‐Multiwalled Carbon Nanotubes Nanocomposite/Bi Film Electrode for Sensitive Determination of Cd and Pb by Differential Pulse Stripping Voltammetry
Authors:Wenfang Deng  Yueming Tan  Zhengfa Fang  Qingji Xie  Yunyong Li  Xiaoling Liang  Shouzhuo Yao
Institution:Key Laboratory of Chemical Biology and Traditional Chinese Medicine Research (Ministry of Education of China), College of Chemistry and Chemical Engineering, Hunan Normal University, Changsha 410081, P.?R. China
Abstract:A 2,2′‐azinobis (3‐ethylbenzothiazoline‐6‐sulfonate) diammonium salt (ABTS)‐multiwalled carbon nanotubes (MWCNTs) nanocomposite/Bi film modified glassy carbon (GC) electrode was constructed for the differential pulse stripping voltammetric determination of trace Pb2+ and Cd2+. This electrode was more sensitive than ABTS‐free Bi/GC and Bi/MWCNTs/GC electrodes. Linear responses were obtained in the range from 0.5 to 35 μg L?1 for Cd2+ and 0.2 to 50 μg L?1 Pb(II), with detection limits of 0.2 μg L?1 for Cd2+ and 0.1 μg L?1 for Pb2+, respectively. This sensor was applied to the simultaneous detection of Cd2+ and Pb2+ in water samples with satisfactory recovery.
Keywords:2  2′  ‐Azinobis(3‐ethylbenzothiazoline‐6‐sulfonate) diammonium salt (ABTS)  Multiwalled carbon nanotubes (MWCNTs)  Bismuth film electrode (BiFE)  Differential pulse stripping voltammetry (DPSV)  Pb2+  Cd2+  Voltammetry  Nanotube