Objectives
A patient with a breast tissue expander may require a diagnostic assessment using magnetic resonance imaging (MRI). To ensure patient safety, this type of implant must undergo in vitro MRI testing using proper techniques. Therefore, this investigation evaluated MRI issues (i.e., magnetic field interactions, heating, and artifacts) at 3-Tesla for a breast tissue expander with a remote port.Methods
A breast tissue expander with a remote port (Integra Breast Tissue Expander, Model 3612-06 with Standard Remote Port, PMT Corporation, Chanhassen, MN) underwent evaluation for magnetic field interactions (translational attraction and torque), MRI-related heating, and artifacts using standardized techniques. Heating was evaluated by placing the implant in a gelled-saline-filled phantom and MRI was performed using a transmit/receive RF body coil at an MR system reported, whole body averaged specific absorption rate of 2.9-W/kg. Artifacts were characterized using T1-weighted and GRE pulse sequences.Results
Magnetic field interactions were not substantial and, thus, will not pose a hazard to a patient in a 3-Tesla or less MRI environment. The highest temperature rise was 1.7 °C, which is physiologically inconsequential. Artifacts were large in relation to the remote port and metal connector of the implant but will only present problems if the MR imaging area of interest is where these components are located.Conclusions
A patient with this breast tissue expander with a remote port may safely undergo MRI at 3-Tesla or less under the conditions used for this investigation. These findings are the first reported at 3-Tesla for a tissue expander. 相似文献2.
用自由基溶液降合方法制备一系列苯乙烯-丙烯酸甲酯线型共聚物,用核磁共振测定了苯基在共聚物中的百分比,在该共聚物的四氢呋喃溶液中,用三氟乙酸汞在共聚物的苯环上进行亲电取代反应,得到可溶性汞化共聚物,由于这类泵化共的可溶于四氢呋喃,二氯甲烷等溶剂,用重沉淀法多次提纯,得到了纯度很高,溶解性较好的含重金属二价汞的共聚物,用红外光谱仪测定共聚物上的汞基团,用原子吸收定量测试共聚中的汞的百分聚代率,结果表明 相似文献
3.
Reductions in MRI-induced heating at 1.5 T resulting from a simple design change to coiled wires were investigated. MRI-induced heating was assessed for two different coiled wire forms (length, 26 cm): (1) multi-filar coiled wire form and (2) multi-filar coiled wire form having a different coiled pitch, providing an air gap spacing between adjacent five-filar coil loops. Each wire had an electrode and was insulated to create a lead, similar to that which would be used for a medical implant. The wire forms were placed in a gelled-saline-filled head/torso phantom and imaged at 1.5 T [whole-body average specific absorption rate (SAR), 1.79 W/kg]. Fluoroptic thermometry probes were used to measure temperatures at the distal ends of the wires. The experiments demonstrated a substantial reduction in MRI-induced heating for the modified wire compared to the unmodified wire (i.e., 10.5 degrees C difference observed in one experiment and 26 degrees C difference in another). These findings have important implications for MRI-induced heating of leads used for medical implants. 相似文献
4.
5.
Background and Purpose
A recent report suggested that a serious burn injury was due to the presence of the identification (ID) wristband. As such, in lieu of removing or padding hospital ID wristbands in all patients prior to magnetic resonance imaging (MRI), testing may be performed to characterize risks for ID wristbands. Therefore, the purpose of this investigation was to evaluate the magnetic field interactions, heating and artifacts at 3 T for a hospital ID wristband.Materials and Methods
Standardized test methods were used to evaluate magnetic field interactions, MRI-related heating, and artifacts at 3 T for a hospital ID wristband.Results
There were no magnetic field interactions. MRI-related heating evaluated at a relatively high, MR system-reported, whole body-averaged specific absorption rate (2.9 W/kg) did not increase above the background level. The artifacts related to the ink used for printing were “small” for one toner and “large” for the other in relation to the dimensions of the printing.Conclusions
Based on the tests performed, this particular hospital ID wristband is considered MR safe and will not pose a hazard to a patient undergoing an MRI examination. Importantly, it is not necessary to remove this item for a patient referred for MRI. 相似文献6.
Objective
Vascular grafting frequently involves a time-consuming operation. A new vascular coupling device (VCD) made from metallic material was recently developed that may be advantageous because of the reduced operative time and decreased patient risks. Because of the metal, there are safety concerns related to MRI. Therefore, the purpose of this investigation was to use standardized testing techniques to evaluate MRI issues for this VCD in association with a 3-Tesla MR system.Methods
The VCD (corlife oHG, Hannover, Germany) was evaluated for magnetic field interactions (translational attraction and torque), MRI-related heating, and artifacts at 3-Tesla. MRI-related heating was assessed with the VCD in a gelled-saline-filled phantom with MRI performed at a whole body averaged SAR of 2.9-W/kg for 15-min. Artifacts were assessed using T1-weighted, spin echo, and gradient echo pulse sequences.Results
The VCD exhibited minor magnetic field interactions and minimal heating (maximum temperature elevation, 1.8 °C). Artifacts were relatively small in relation to the size and shape of this implant. The lumen of the VCD could not be visualized using the gradient echo pulse sequence.Conclusions
The metallic VCD that underwent evaluation is MR conditional for a patient undergoing an MRI procedure at 3-Tesla or less. 相似文献7.
Christina L. Sammet Xiangyu Yang Peter A. Wassenaar Eric C. Bourekas Brian A. Yuh Frank Shellock Steffen Sammet Michael V. Knopp 《Magnetic resonance imaging》2013
Purpose
The purpose was to evaluate radiofrequency (RF)-related heating of commonly used extracranial neurosurgical implants in 7-T magnetic resonance imaging (MRI).Materials and methods
Experiments were performed using a 7-T MR system equipped with a transmit/receive RF head coil. Four commonly used titanium neurosurgical implants were studied using a test procedure adapted from the American Society for Testing and Materials Standard F2182-11a. Implants (n = 4) were tested with an MRI turbo spin echo pulse sequence designed to achieve maximum RF exposure [specific absorption rate (SAR) level = 9.9 W/kg], which was further validated by performing calorimetry. Maximum temperature increases near each implant's surface were measured using fiberoptic temperature probes in a gelled-saline-filled phantom that mimicked the conductive properties of soft tissue. Measurement results were compared to literature data for patient safety.Results
The highest achievable phantom averaged SAR was determined by calorimetry to be 2.0 ± 0.1 W/kg due to the highly conservative SAR estimation model used by this 7-T MR system. The maximum temperature increase at this SAR level was below 1.0 °C for all extracranial neurosurgical implants that underwent testing.Conclusion
The findings indicated that RF-related heating under the conditions used in this investigation is not a significant safety concern for patients with the particular extracranial neurosurgical implants evaluated in this study. 相似文献8.
Purpose
To evaluate MRI artifacts at 3-Tesla for 38 commonly used cosmetics.Materials and Methods
Thirty-eight cosmetics (16, nail polishes; 5, eyeliners; 3, mascaras; 10, eye shadows; 1, lip gloss; 1, body lotion; 1, body glitter, and 1, hair loss concealer) underwent evaluation for MRI artifacts at 3-Tesla. The cosmetics were applied a copper-sulfate-filled, phantom and initially assessed using a “screening” gradient echo (GRE) pulse sequence. Of the 38 different cosmetics, 14 (37%) exhibited artifacts. For these 14 cosmetics, additional characterization of artifacts was performed using a GRE pulse sequence. A qualitative scale was applied to characterize the artifact size.Results
Artifacts were observed, as follows: 2, nail polishes; 5, eyeliners; 3, mascaras; 3, eye shadows; 1, hair loss concealer. Artifact size ranged from small (eye shadow) to very large (hair loss concealer) and tended to be associated with the presence of iron oxide or other metal-based ingredient.Conclusions
Commonly used cosmetics caused artifacts that may create issues if the area of interest is the same as where the cosmetic was applied or if its presence was unknown, thus, potentially causing it to be construed as pathology. Therefore, these findings have important implications for patients referred for MRI examinations. 相似文献9.
10.
Weiland JD Faraji B Greenberg RJ Humayun MS Shellock FG 《Magnetic resonance imaging》2012,30(3):382-389