Gold‐functionalized stainless‐steel wire and tube for fiber‐in‐tube solid‐phase microextraction coupled to high‐performance liquid chromatography for the determination of polycyclic aromatic hydrocarbons
Authors:Yanan Bu  Juanjuan Feng  Min Sun  Changli Zhou  Chuannan Luo
Affiliation:Key Laboratory of Chemical Sensing & Analysis in Universities of Shandong (University of Jinan), School of Chemistry and Chemical Engineering, University of Jinan, Jinan, China
Abstract:A fiber‐in‐tube solid‐phase microextraction device based on a gold‐functionalized stainless‐steel wire and tube was developed and characterized by scanning electron microscopy and energy dispersive X‐ray spectroscopy. In combination with high‐performance liquid chromatography, it was evaluated using six polycyclic aromatic hydrocarbons as model analytes. Important parameters including sampling rate, sample volume, organic solvent content and desorption time were investigated. Under optimized conditions, an online analysis method was established. The linearity was in the range of 0.15–50 μg/L with correlation coefficients ranging from 0.9989 to 0.9999, and limits of detection ranged from 0.05 to 0.1 μg/L. The method was applied to determine model analytes in mosquito‐repellent incense ash and river water samples, with recoveries in the range of 85–120%.
Keywords:Fiber‐in‐tube  Gold  Polycyclic aromatic hydrocarbons  Solid‐phase microextraction  Stainless steel