首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Status of the front-end-electronics for the time-of-flight measurements at the MPD experiment
Authors:M G Buryakov  V A Babkin  V M Golovatyuk  S V Volgin  M M Rumyantsev
Institution:1.Joint Institute for Nuclear Research,Dubna,Russia
Abstract:The preamplifier based on the ASIC NINO for the Time of Flight system (TOF) of MPD/NICA was developed and tested. The signal is read from both sides of the strip of the multi gap Resistive Plate Chamber (MRPC). In total there are around 14000 channels of electronics. To measure time of flight of secondary particles from collision of heavy ions on the collider NICA the Time over Threshold (ToT) method is used. According to the bench tests the preamplifier board showed stable work and good time resolution <10 ps for one channel. It was also tested at the test beam facility of the Nuclotron. The time resolution of the TOF detector which used the described preamplifier was reached ~42 ps.