Direct determination of benzylnitriles using the cyanide ion selective electrode |
| |
Authors: | Saad S. M. Hassan M. A. Mari |
| |
Affiliation: | (1) Department of Chemistry, Faculty of Science, Ain Shams University, Egypt;(2) Microanalytical Research Centre, Cairo University, Egypt;(3) Department of Chemistry, Faculty of Science, Qatar University, P. O. Box 2713, Doha, Qatar |
| |
Abstract: | The response characteristics of the solid state cyanide ion selective electrode towards some benzylnitriles are investigated. In 10M KOH solution, the electrode exhibits nearly Nernstian response over the concentration range of 10–2 to 10–4M of various substituted benzylnitriles with an anionic slope of 53–59 mV/concentration decade. The response time varies from 10 to 20 min depending on both the nature of the substituent group and the concentration of the nitrile compound. Direct potentiometric measurement of some nitrile compounds at the concentration level of 0.01 to 1 mg/ml shows an average recovery of 98.2% and a mean standard deviation of 2.3%. Many nitrogen functional groups do not interfere. |
| |
Keywords: | potentiometry cyanide electrode determination of benzylnitriles |
本文献已被 SpringerLink 等数据库收录! |
|