首页 | 本学科首页   官方微博 | 高级检索  
     


Suppression correction and characteristic study in liquid chromatography/Fourier transform mass spectrometry measurements
Authors:Ma Xuepo  Hestilow Travis J  Cui Jian  Zhang Jianqiu
Affiliation:Department of Electrical and Computer Engineering, University of Texas at San Antonio, San Antonio, TX, USA.
Abstract:Analysis of peptide profiles from liquid chromatography/Fourier transform mass spectrometry (LC/FTMS) reveals a nonlinear distortion in intensity. Investigation of the measured C(13)/C(12) ratios comparing with theoretical ones shows that the nonlinearity can be attributed to signal suppression of low abundance peptide peaks. We find that the suppression is homogenous for different isotopes of identical peptides but non-homogenous for different peptides. We develop an iterative correction algorithm that corrects the intensity distortions for peptides with relatively high abundance. This algorithm can be applied in a wide range of applications using LC/FTMS. We also analyze the distortion characteristics of the instrument for lower abundance peptides, which should be considered when interpreting quantification results of LC/FTMS.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号