Long time stability of the energy scale calibration of a Quantum 2000 |
| |
Authors: | Uwe Scheithauer |
| |
Affiliation: | SIEMENS AG, CT T DE HW1, Otto-Hahn-Ring 6, 81739 München, Germany |
| |
Abstract: | According to the international standard ISO 15472 the energy scale of an XPS instrument, type Physical Electronics Quantum 2000, was calibrated. It is shown, how the procedures of the ISO 15472 were adapted to the hardware and software design of the Quantum 2000. The long time stability of the energy scale calibration of the XPS instrument was investigated. The instrumented was operated with a satisfying energy scale calibration over a period of 8 years. All the time energy differences between certain peaks could be measured with the chosen precision of the energy scale calibration. |
| |