首页 | 本学科首页   官方微博 | 高级检索  
     


Deactivation of small-diameter fused-silica capillary columns for gas and supercritical fluid chromatography
Authors:R. C. Kong  C. L. Woolley  S. M. Fields  M. L. Lee
Affiliation:(1) Department of Chemistry, Brigham Young University, 84602 Provo, UT, USA
Abstract:Summary In order to prevent plugging during deactivation of small diameter (50 mgrm i.d.) capillary columns for gas and supercritical fluid chromatography, various high temperature deactivation methods were employed. Pure hexamethyldisilazane and hexamethyldisiloxane (a substitute for D4) were dynamically coated on the column, while a film (0.05 mgrm) of OV-101 was statically coated, before high temperature (450°C) treatment. Excellent deactivated columns were obtained, and no significant difference in column activity was observed using any of these three methods.Dedicated to Professor S. R. Lipsky on the occasion of his 60th birthday.
Keywords:Deactivation of columns  Capillary gas chromatography  Capillary supercritical fluid chromatography
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号