首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of a condensation nucleation light scattering detector for the analysis of synthetic polymer by supercritical fluid chromatography
Authors:Kayori Takahashi  Shinichi Kinugasa  Ryo Yoshihara  Akira Nakanishi  Renee K Mosing  Rheo Takahashi
Institution:1. National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba, Ibaraki, Japan;2. Asahi Techneion Co., LTD., Tokyo, Japan;3. Quant Technologies LLC, Blaine, MN, USA;4. Department of Chemistry and Chemical Biology, Graduate School of Engineering, Gunma University, 1-5-1 Tenjin, Kiryu, Gunma 376-8515, Japan
Abstract:A condensation nucleation light scattering detector (CNLSD) was adapted for use as a detector for supercritical fluid chromatography. The performance of the CNLSD was evaluated and compared to evaporative light-scattering detector (ELSD) using a well-defined equimass mixture of uniform poly(ethylene glycol) oligomers and a certified reference material of poly(ethylene glycol) 1000. The CNLSD was able to detect a 10 times less concentrated solution of uniform oligomers compared to the ELSD. The quantitativeness of CNLSD was high enough to obtain the molecular mass distribution of poly(ethylene glycol) 1000 without any calibrations; on the other hand, the original data measured by ELSD was about 4% smaller than the certified value of poly(ethylene glycol) 1000. The CNLSD was suitable for supercritical fluid chromatography as a mass concentration detector.
Keywords:CNLSD  NQAD  ELSD  SFC  Poly(ethylene glycol)  Calibration
本文献已被 ScienceDirect 等数据库收录!