首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Effects of pressure drop,particle size and thermal conditions on retention and efficiency in supercritical fluid chromatography
Authors:Donald P Poe  Jonathan J Schroden
Institution:Department of Chemistry and Biochemistry, University of Minnesota Duluth, 10 University Drive, Duluth, MN 55812, USA
Abstract:The effects of particle size and thermal insulation on retention and efficiency in packed-column supercritical fluid chromatography with large pressure drops are described for the separation of a series of model n-alkane solutes. The columns were 2.0 mm i.d. × 150 mm long and were packed with 3, 5, or 10-μm porous octylsilica particles. Separations were performed with pure carbon dioxide at 50 °C at average mobile phase densities of 0.47 g/mL (107 bar) and 0.70 g/mL (151 bar). The three principal causes of band broadening were the normal dispersion processes described by the van Deemter equation, changes in the retention factor due to the axial density gradient, and radial temperature gradients associated with expansion of the mobile phase. At the lower density the use of thermal insulation resulted in significant improvements in efficiency and decreased retention times at large pressure drops. The effects are attributed to the elimination of radial temperature gradients and the concurrent enhancement of the axial temperature gradient. Thermal insulation had no significant effect on chromatographic performance at the higher density. A simple expression to predict the onset of excess efficiency loss due to the radial temperature gradient is proposed.
Keywords:Supercritical fluid chromatography  Pressure drop  Efficiency  Retention  Temperature gradients  Thermal conditions  Modeling
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号