MICROMEGAS探测器初步测试
引用本文:鲁辰桂,段利敏,杨贺润,马朋,唐述文,耿朋,胡荣江,张金霞,李祖玉.MICROMEGAS探测器初步测试[J].原子核物理评论,2011,28(1):68-70.
作者姓名:鲁辰桂  段利敏  杨贺润  马朋  唐述文  耿朋  胡荣江  张金霞  李祖玉
作者单位:1 中国科学院近代物理研究所, 甘肃 兰州 730000; 2 中国科学院研究生院, 北京 100049
基金项目:国家自然科学基金资助项目,中国科学院知识创新工程重大项目,国家重点基础研究发展计划
摘    要:介绍了一种利用鱼线做放大区间隔的MICROMEGAS探测器, 灵敏面积为50 mm ×50 mm。 在Ar+CO2(10%)气体条件下, 利用55Fe源5.9 keV的 X射线对MICROMEGAS探测器进行了初步测试: 计数率大于103 Hz的条件下, 计数率坪长达到280 V; 当栅极电压在800 V时能量分辨为30%; 当增益高于104时打火率低于10-4; PCB板读出条周期为400 μm条件下的位置分辨好于120 μm。 结果表明, 研制的MICROMEGAS探测器达到了初步设计的基本要求。 In this paper,A MICROMEGAS(Micro Mesh Gaseous Structure) detector with the amplification gap defined by stretching nylon fishing lines has been described . And its count plateau, gas gain and energy resolution in the condition of a 55Fe X ray source and Ar CO2 (10%) mixture gases are measured as functions of mesh voltage. The count plateau approaches 280 V at the count rate of 103 Hz. The discharge rate is less than 10-4 when the gas gain is up to 104. The energy resolution is 30% (FWHM) based on -800 V mesh voltage. The position resolution is less than 120 μm at the condition of the sensitive area (50 mm×50 mm) Printed Circuit Boards (PCB) and its strip pitch of 400 μm. The results can satisfy the basic demand of MICROMEGAS detector preliminary design.

关 键 词:计数率坪曲线    能量分辨    增益    打火率    位置分辨
收稿时间:1900-01-01

Preliminary Test of MICROMEGAS Gaseous Detector
LU Chen-gui,DUAN Li-min,YANG He-run,MA Peng,TANG Shu-wen,GENG Peng,HU Rong-jiang,ZHANG Jin-xia,LI Zu-yu.Preliminary Test of MICROMEGAS Gaseous Detector[J].Nuclear Physics Review,2011,28(1):68-70.
Authors:LU Chen-gui  DUAN Li-min  YANG He-run  MA Peng  TANG Shu-wen  GENG Peng  HU Rong-jiang  ZHANG Jin-xia  LI Zu-yu
Institution:1 Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China;2 Graduate University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:In this paper,A MICROMEGAS(Micro Mesh Gaseous Structure) detector with the amplification gap defined by stretching nylon fishing lines has been described . And its count plateau, gas gain and energy resolution in the condition of a 55Fe X ray source and Ar CO2 (10%) mixture gases are measured as functions of mesh voltage. The count plateau approaches 280 V at the count rate of 103 Hz. The discharge rate is less than 10-4 when the gas gain is up to 104. The energy resolution is 30% (FWHM) based on -800 V mesh voltage. The position resolution is less than 120 μm at the condition of the sensitive area (50 mm×50 mm) Printed Circuit Boards (PCB) and its strip pitch of 400 μm. The results can satisfy the basic demand of MICROMEGAS detector preliminary design.
Keywords: