首页 | 本学科首页   官方微博 | 高级检索  
     


Cyclodextrin-MEEKC for the analysis of oxybutynin and its impurities
Authors:Iacopo Giannini  Roberto Gotti  Sandra Furlanetto
Affiliation:a Department of Pharmaceutical Sciences, University of Florence, Via U. Schiff 6, 50019 Sesto Fiorentino, Florence, Italy
b Department of Pharmaceutical Sciences, University of Bologna, Via Belmeloro 6, 40126 Bologna, Italy
Abstract:The development of a cyclodextrin-MEEKC method for the analysis of oxybutynin and five related impurities is described. Experimental design strategies were applied in order to reach baseline separation of the compounds in a short analysis time. Mixture design made it possible to find the best composition for the microemulsion acting as pseudostationary phase, which was constituted by 89.1% 10 mM borate buffer pH 9.2, 1.7% n-heptane, 9.2% SDS/n-butanol in 1:2 ratio. The addition of (2-hydroxypropyl)-β-cyclodextrin to the background electrolyte was found to improve analysis performance. A Doehlert design, for the factors cyclodextrin concentration and voltage, was carried out and Derringer desirability function led to the identification of 18 mM and 29 kV as the optimal values. Applying the optimum conditions, separation of all the compounds, including the enantiomers of impurity 1, was obtained in less than 12 min. The method was validated according to ICH guidelines for drug assay and determination of impurities and was applied to oxybutynin tablet analysis.
Keywords:Cyclodextrin   Doehlert design   Impurities   MEEKC   Mixture design   Oxybutynin
本文献已被 ScienceDirect 等数据库收录!