Abstract:The traditional method for purity determination by analysis of the peak shape of the melting transition has included fitting to a linear relation between sample temperature (T) and the reciprocal fraction molten (17). This technique, however, necessitates the use of a series of calibrants to determine the proper choice of limits for 17 as a function of purity. In this paper, a non-linear relation between T and 17 is developed and applied to the determination of metallic impurities in Pb in the range of a few to 1000 parts per million. The results are found to be independent of the range of 17 used. The use of differential scanning calorimety in this determination is discussed, and rate effects are also mentioned.