Assessments on energy and efficiency calibration of an alpha spectrometry system using standard sources
Authors:M R Calin  M A Saizu  I Radulescu
Institution:1. Department of Life and Environmental Physics, “Horia Hulubei” National Institute for Physics and Nuclear Engineering—IFIN HH, 30 Reactorului Street, P.O. Box MG-6, 077125, Magurele, Romania
Abstract:Studies and evaluations of the main parameters of a high resolution multi-chamber alpha spectrometer and its full calibration using standard sources of 241Am, 233U and 244Cm are presented for the validation of the complex method of characterization of an alpha spectrometer dedicated for alpha radioactive concentration measurements of various environmental samples. The resolution and efficiency were determined at all possible source-detector distances for all eight solid state ORTEC ULTRA-AS detectors. In addition, the solid angles associated to different measurement geometries and the repeatability of the results were assessed. A detailed inter-comparison of the results was performed drawing important conclusions regarding the quality of the alpha detector response to the alpha radiation of the standard sources used in measurements.