Air handling in clean laboratory environments: the reason for anomalously high boron background levels
Authors:Email author" target="_blank">Martin?RosnerEmail author  Rolf?L?Romer  Anette?Meixner
Institution:(1) GeoForschungsZentrum Potsdam, 14473 Potsdam, Germany
Abstract:The isotopic analysis of boron from material that carries only trace amounts of boron requires low boron blank levels. Large efforts are taken to keep blank levels low by purifying reagents. We performed exposure experiments and determined procedure blanks before and after the air-handling system was modified from silicate glass filters to polyethylene/polystyrole and active carbon filters. Our investigations demonstrate that the air-handling system may be the major source of boron blank in many laboratories, as the widely used silicate glass filters, which contain boron as a major component, in the form of borax and boric acid, release significant amounts of boron to the air.
Keywords:Boron contamination  Boron blank  Volatile boron  Filter materials  Clean laboratory facilities  Air-handling systems
本文献已被 PubMed SpringerLink 等数据库收录!