首页 | 本学科首页   官方微博 | 高级检索  
     

气相色谱中测定死时间的Grobler-Balizs法和Ambrus法的比较
引用本文:吴宁生,苏红伟,史文娟. 气相色谱中测定死时间的Grobler-Balizs法和Ambrus法的比较[J]. 色谱, 1996, 0(1)
作者姓名:吴宁生  苏红伟  史文娟
作者单位:中国科技大学应用化学系
摘    要:经理论推导和实验数据验证,证明测定死时间的Grobler-Balizs法和Ambrus法是同一的。

关 键 词:气相色谱法,死时间,线性回归

Comparison of Grobler-Balizs Method with Ambrus Method for the Determination of Dead Time in Gas Chromatography
Wu Ningsheng,Su Hongwei and Shi Wenjuan. Comparison of Grobler-Balizs Method with Ambrus Method for the Determination of Dead Time in Gas Chromatography[J]. Chinese journal of chromatography, 1996, 0(1)
Authors:Wu Ningsheng  Su Hongwei  Shi Wenjuan
Abstract:t has been revealed that Grobler-Balizs method and Ambrus method used for determination of dead time in gas chromatography are essentially identical. It has been proved theoretically that the dead time calculated from either method can be expressed as :(qtR.z- tR.z 1 )/(q-1 ).Where Z is the carbon number and q is the ratio of the adjusted retention times of sucessive members of the homologous series,tR is the retention time.The calculated results of experimental data reported in one refererce (Wronski B et al.J.Chromatogr,1986;364:53) showed that the limits of errors between these two methods ranged from- 2.05% to 0. 68%.The error is caused by different regression programs used in these two methods.Grobler-Balizs method needs two linear regression programs but Ambrus method needs only one.The latter is simpler.
Keywords:gas chromatography   dead time   linear regression  
本文献已被 CNKI 等数据库收录!