Quantitative determination of propranolol in plasma and plasma water from normal subjects and patients with angina pectoris by high-performance liquid chromatography |
| |
Authors: | Y Yamamura K Uchino H Kotaki S Isozaki Y Saitoh |
| |
Abstract: | A precise and sensitive high-performance liquid chromatographic method using a column packed with porous polystyrene gel is described for the determination of propranolol in plasma and plasma water from normal subjects and patients with angina pectoris. Propranolol in the samples was extracted with an n-heptane-isoamylalcohol (98.5:1.5) mixture after addition of penbutolol used as an internal standard. The extracts were chromatographed and detected with a spectrofluorophotometer. The quantitative limit of propranolol was 1 ng using 1 ml of plasma or 0.5 ml of plasma water. The present method should be useful for monitoring propranolol concentrations in plasma and plasma water during drug therapy and for pharmacokinetic study of propranolol. |
| |
Keywords: | |
|