首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Capacitive Chemical Sensor for Thiopental Assay Based on Electropolymerized Molecularly Imprinted Polymer
Authors:Mostafa Najafi  Amin Azam Baghbanan
Institution:Department of Chemistry, Faculty of Science, Imam Hossein University, Tehran, 16597, Iran
Abstract:A novel capacitive sensor based on electropolymerized molecularly imprinted polymer (MIP) for thiopental detection is described. The molecularly imprinted film as a recognition element was prepared by electropolymerization of phenol on a gold electrode in the presence of thiopental (template). Cyclic voltammetry and capacitive measurements were used for characterization and evaluation of the polymeric film. The template molecules were removed from the modified electrode surface by washing with an ethanol:water solution. The sensor’s linear response range was between 3 and 20 µM, with a detection limit of 0.6 µM. The proposed sensor exhibited good selectivity, reproducibility. Satisfactory results were obtained in the direct detection of real samples.
Keywords:Capacitive sensors  Molecularly imprinted polymer  Thiopental  Electrochemical impedance spectroscopy
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号