首页 | 本学科首页   官方微博 | 高级检索  
     


Sensitive and rapid determination of nitric oxide in human serum using microchip capillary electrophoresis with laser-induced fluorescence detection
Authors:Yan Wang  Ming Yin
Affiliation:1. College of Chemistry, Chemical Engineering and Materials Science, Shandong Normal University, Jinan, 250014, China
2. Institute of Electron Beam, School of Control Science and Engineering, Shandong University, Jinan, 250061, China
Abstract:Microfluidic chip capillary electrophoresis with laser-induced fluorescence detection is employed for direct determination of trace nitric oxide in human blood using diaminorhodamines as the fluorescence probe. Factors influencing the separation and detection processes were systematically studied. Complete and fast separation of the highly fluorescent triazole formed was achieved within 45 s, and the relative standard deviations values of migration time and peak area were less than 3%. The detection limit of NO was 3.0 nmol.L-1 (at a signal-to-noise ratio of 3) and the liner range was from 1.0?×?10-8?mol.L-1 to 3.0?×?10-6?mol.L-1. The method has been applied to the determination of NO in serum of healthy persons and patients suffering from diseases, with recoveries varying from 92.65 to 98.43%.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号