Comparison of matrix effects in inductively coupled plasma using laser ablation and solution nebulization for dry and wet plasma conditions
Institution:1. Department of Chemistry, The University of Hong Kong, Pokfulam Road, Hong Kong;2. Lawrence Berkeley National Laboratory, 1 Cyclotron Road, Berkeley, CA 94720, USA;1. Leica Geosystems Technology A/S, Telehøjen 8, DK-5220 Odense, Denmark;2. Institute of Photogrammetry and GeoInformation, Leibniz Universität Hannover, Nienburger Str. 1, D-30167 Hannover, Germany;3. University of Twente, Hengelosestraat 99, Enschede, The Netherlands
Abstract:Matrix effects of calcium in inductively coupled plasma-atomic emission spectrometry were investigated. Matrix effects were studied by monitoring the excitation conditions of the plasma using Zn ionic to atomic spectral line intensity ratios. Dry and wet inductively coupled plasmas with robust and non-robust conditions were compared. Laser ablation and solution nebulization sample introduction were used to produce the dry and wet plasma conditions, respectively. Low (0.6 l/min) and high (1.0 l/min) carrier gas flow rates were used to produce the robust and non-robust conditions, respectively. No differences in the trend of matrix effects for dry and wet plasmas were observed at vertical positions above normal observation height (>8 mm height above load coil) for low and high carrier gas flow rates. However, matrix effects in the lower part of the plasma (<8 mm height above load coil) were significantly different between dry and wet plasmas when a high carrier gas flow was used. The differences are likely due to the desolvation process.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号