首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Lab-on-valve for the automatic determination of the total content and individual profiles of linear alkylbenzene sulfonates in water samples
Authors:Ruiz Jiménez José  Luque de Castro María D
Institution:Department of Analytical Chemistry, Annex Marie Curie Building. Campus of Rabanales, University of Córdoba, Córdoba, Spain.
Abstract:Methods for the total content and individual determination of linear alkylbenzene sulfonates (LAS) in water samples based on the use of a lab-on-valve (LOV) module alone or coupled to CE equipment, respectively, have been developed. The total content of LAS has been determined by intrinsic absorption measurements (DA method) and after reaction with a methyl orange-cetylpyridine chloride mixture (MO method) with detection limits (LODs) of 21 ng/L and 15 microg/L, respectively, quantification limits (LOQs) of 70 ng/L, 50 microg/L, and development times of 100 and 124 s, respectively. The method for individual separation-quantification of LAS at very low concentration is based on automatic SPE preconcentration in the LOV module coupled on-line with the CE equipment. The LODs and LOQs thus obtained range between 1-21 and 4-70 ng/L, respectively, with linear dynamic ranges from the LOQ to 10 microg/L. Preconcentration factors of 10,000 and high efficiency to eliminate interferents by SPE enable application of the method to treated effluent, waste, surface and sea waters.
Keywords:Capillary electrophoresis  Direct absorbance method  Lab‐on‐valve  Linear alkylbenzene sulphonate  Methyl orange
本文献已被 PubMed 等数据库收录!