Slow equilibration of reversed-phase columns for the separation of ionized solutes
Authors:Marchand D H  Williams L A  Dolan J W  Snyder L R
Affiliation:Department of Chemistry, University of Wisconsin River Falls, River Falls, WI 54022, USA. daniel.marchand@uwrf.edu
Abstract:Reversed-phase columns that have been stored in buffer-free solvents can exhibit pronounced retention-time drift when buffered, low-pH mobile phases are used with ionized solutes. Whereas non-ionized compounds exhibit constant retention times within 20 min of the beginning of mobile phase flow, the retention of ionized compounds can continue to change (by 20% or more) for several hours. If mobile phase pH is changed from low to high and back again, an even longer time may be required before the column reaches equilibration at low pH. The speed of column equilibration for ionized solutes can vary significantly among different reversed-phase columns and is not affected by flow rate.