Separation of chromosomal length DNA molecules: pneumatic apparatus for rotating gels during electrophoresis
Authors:J C Sutherland  A B Emrick  J Trunk
Institution:Biology Department, Brookhaven National Laboratory, Upton, NY 11973.
Abstract:The maximum length of DNA molecules that can be separated by gel electrophoresis can be increased greatly by periodically altering the direction of the electric field with respect to the gel by an angle that exceeds 90 degrees. One method involves rotating the gel by the desired angle in alternate directions periodically during electrophoresis. We describe a modification of the rotating gel electrophoresis apparatus developed by Serwer (Electrophoresis 1987, 8, 301-304) that uses a pneumatic rotary actuator instead of a stepping motor, hence reducing the cost by about 50%. Other advantages of our design are a lower center of gravity that makes the apparatus more stable and the removal of all electrical power from beneath the fluid-filled electrophoresis chamber. We present data demonstrating the separation of chromosomal length DNA molecules from Saccharomyces cerevisiae strain 334 into 14 resolved bands in parallel lanes.