Suppressed electrostatic ion chromatography with tetraborate as eluent and its application to the determination of inorganic anions in snow and rainwater
Authors:Hu W  Tanaka K  Haddad P R  Hasebe K
Institution:Division of Chemistry. Graduate School of Science, Hokkaido University, Sapporo, Japan.
Abstract:Tetraborate is investigated as the eluent ion for suppressed electrostatic ion chromatography (EIC) using a zwitterionic stationary phase. Good separation of a range of inorganic anions (SO4(2-), Cl-, NO3-, Br-, NO3-, ClO3-, and I-) was obtained, with detection limits for highly conducting ions (SO4(2-), Cl-, NO2-, Br- , and NO3-) being less than 8 x 10(-8) M, and for weakly conducting anions (ClO3- and I-) being 2.7 x 10(-7) and 5.8 x 10(-7) M, respectively. Calibration curves were linear up to 1.8 mM of each analyte. Retention times were found to increase with increasing eluent concentration and a linear relationship was observed between log k' and logNa2B4O7] for all analytes. This behaviour is attributed to the progressive formation of a binary electrical double layer at the surface of the zwitterionic stationary phase. Retention times of analytes could be manipulated by varying the concentration of the eluent. This new suppressed-EIC system was applied to the determination of inorganic anions (SO4(-2) , CI-, NO3-, NO2-, and Br-) in snow and rainwater samples.