首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Validation of a stability-indicating RP-LC method for the assessment of recombinant human interleukin-11 and its correlation with bioassay
Authors:Souto Ricardo Bizogne  Stamm Fernanda Pavani  Ribela Maria Teresa de Carvalho Pinto  Bartolini Paolo  Calegari Guilherme Zanini  Dalmora Sérgio Luiz
Institution:Department of Industrial Pharmacy, Federal University of Santa Maria, Santa Maria-RS, Brazil.
Abstract:A stability-indicating reversed-phase liquid chromatography (RP-LC) method was validated for the assessment of recombinant human interleukin-11 (rhIL-11), based on the ICH guidelines. The method was carried out on a Jupiter C(4) column (250 mm × 4.6 mm i.d.), maintained at 25°C. The mobile phase A consisted of 0.1% trifluoroacetic acid (TFA) and the mobile phase B was acetonitrile with 0.1% TFA, run at a flow rate of 1 mL/min, and using a photodiode array (PDA) detection at 214 nm. Separation was obtained with a retention time of 27.6 min, and was linear over the concentration range of 1-200 μg/mL (r(2) = 0.9995). Specificity was established in degradation studies, which also showed that there was no interference of the excipients. The accuracy was 100.22% with bias lower than 1.25%. Moreover, the in vitro cytotoxicity test of the degraded products showed non-significant differences (p > 0.05). The method was applied to the assessment of rhIL-11 and related proteins in biopharmaceutical dosage forms, and the results were correlated to those of a bioassay.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏