首页 | 本学科首页   官方微博 | 高级检索  
     


Advantages of low-background liquid scintillation alpha-spectrometry and pulse shape analysis in measuring222Rn, uranium and226Ra in groundwater samples
Authors:L. Salonen  H. Hukkanen
Affiliation:(1) Finnish Centre for Radiation and Nuclear Safety, P.O. Box 14, Fin-00881 Helsinki, Finland
Abstract:Liquid scintillation counting (LSC) and pulse shape analysis (PSA) was used in measuring radon and gross alpha- and beta-activities in groundwater. We used conventional LSC counters for the measurement of radon in water, but low-background LSC spectrometers for the gross activity measurements. The lower limit of detection (LLD) for radon in water is 0.6 Bq/l for a 60 min count with a conventional counter, but 0.1 or 0.2 Bq/l, with the two types of low-background LSC spectrometers equipped with a pulse shape analyser (PSA). The gross alpha and beta activity measurements are made using a simple sample preparation method, PSA of a low background LSC and spectrum analysis. The LLD recorded for gross alpha and beta with the two spectrometers are 0.02 and 0.03 Bq/l and 0.2 and 0.4 Bq/l, respectively, for a 180 minutes count and a 38 ml sample volume. The method also enable the calculation of the U and226Ra contents in water and indicates the presence of some other long-lived radionuclides (210Pb,228Ra or40K). The LLD for U recorded with both spectrometers is 0.02 Bq–1 and for226Ra 0.01 Bq·1–1. The LLDs attained by this LSC method are two orders of magnitude lower than the maximum permissible concentrations set for U and226Ra.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号