Inter-laboratory performance between two nanoparticle air filtration systems using scanning mobility particle analyzers
Authors:Michael B Lore  Anthony R Sambol  Daniel A Japuntich  Luke M Franklin  Steven H Hinrichs
Institution:(1) Department of Pathology and Microbiology, University of Nebraska Medical Center, Omaha, NE 68198-6495, USA;(2) 3M Company, Building 201-2E-04, Saint Paul, MN 55144, USA;(3) University of Minnesota, 1100 Mechanical Engineering, 111 Church Street SE, Minneapolis, MN 55455-0111, USA;
Abstract:The performance of two aerosol testing systems, at two different laboratories (University of Nebraska Medical Center—UNMC and 3M Company), was compared to evaluate which calibration procedures minimized variability in filter testing of nanoparticles. Both charged electret and uncharged flat-web fibrous filters were used with Scanning Mobility Particle Sizers to give upstream and downstream size distributions and calculate filter penetration. Challenge aerosols were polydisperse nanoparticles of sodium chloride (NaCl) ranging from approximately 10–300 nm and monodisperse polystyrene latex (PSL) spheres of preselected sizes, including 40, 60, 100, and 200 nm. The implementation of optimized procedures resulted in comparable filtration performance at the two testing sites with challenges of NaCl particles and PSL spheres. The penetration results for the uncharged filter were nearly identical for both challenges, while lower penetration through the charged filter was observed with NaCl aerosol, probably due to differences in NaCl and PSL dielectric constants. Results showed that reproducible, comparable nanoparticle filtration data could be achieved between two separate laboratories when sources of error and proper calibration procedures were addressed.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号