A mathematical model based on the limit dilution method to obtain linear calibration curves which eliminate the matrix effect in quantitative analysis by X-ray fluorescence
Institution:1. Microbiología (BIOCLIN), Udelar, Montevideo, Uruguay;2. Química Inorgánica (DEC), Udelar, Montevideo, Uruguay;1. ICAR-Directorate of Medicinal and Aromatic Plants Research, Boriavi, 387 310 Anand, Gujarat, India;2. ICAR-Indian Institute of Horticultural Research, Bengaluru, 560 089, Karnataka, India;3. Horticulture Research Station, Lam, 522 034 Guntur, Andhra Pradesh, India;4. PJ Telangana State Agricultural University, Hyderabad, 500 030, Telangana, India;1. State Key Laboratory of Toxicology and Medical Countermeasures, Laboratory of Toxicant Analysis, Institute of Pharmacology and Toxicology, Academy of Military Medical Sciences, Beijing, 100850, China;2. Key Laboratory of Chemical Biology and Traditional Chinese Medicine Research (Ministry of Education of China), College of Chemistry and Chemical Engineering, Hunan Normal University, Changsha, 410081, China;3. Research Institute of Chemical Defence, Beijing, 100191, China;1. Department of Chemistry, College of Sciences, Taibah University, Al-Medina Al-Munawara 41477, Saudi Arabia;2. Biology Department, Faculty of Sciences, Taibah University, Al-Medina Al-Munawara 41477, Saudi Arabia;3. Department of Chemistry, College of Science, King Saud University, Riyadh 11451, Saudi Arabia;4. Department of Chemistry, Jamia Millia Islamia, (Central University) New Delhi, India
Abstract:We propose a mathematical model from an analytical application viewpoint inspired in the limit dilution method. The theoretical development of the model and its results are given. The model shows that there is a linear relation between the inverse of fluorescence intensity and the inverse of the dilution factor; each analytic system (sample, diluent and analyte) is characterised by a general linear function which is easily obtained. The analytical applications arising from this linearity are of great importance in X-ray fluorescence analysis. The following immediate applications are proposed: direct procurement of the total correction factor Y/H, rapid calculation of the fluorescence intensity of the analyte in a sample (Iis) and direct calculation of the corrected fluorescence intensity (IisF). The suggested model makes it possible to deduce a linear function between the fluorescence intensity of the analyte and the analyte concentration in successive dilutions of a standard; this straight line behaves as a calibration curve with direct application in X-ray fluorescence analysis. The proposed model may be applied to complex samples of geological origin, with elimination of the matrix effect. The results obtained in the determination of Ca, K, Fe and Ti in a standard soil show complete agreement with the certified reference values with a relative error about 0.5%, even using a standard shale with very different chemical composition as reference sample.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号