Abstract:The present paper describes the ashing and atomization processes in silicon analysis by electrothermally atomised atomic absorption spectrometry(EAAAS) with an uncoat-ed graphite tube, a pyrolytically coated graphite tube and a tungsten-coated graphitetube. The sensitivity and linear range of three graphite tubes were compared. By using optical temperature control accessory, the signals are enhanced by a factor of 2 and the germanium interferences in the determination of silicon are eliminated. The effects of time constant and carrier gas flow-rate on the determination of silicon were also tested. The sample can be directly analyzed in its aqueous solution without any pretreatment. The measurements of samples containing 0. 2 μg/mL and 0. 4 μg/mL silicon were run ten times and the variation coefficient is 4. 9% and 2.6%, respectively. The recovery tests for carboxyethyl germanium sesquioxide(Ge-132) synthesized and imported were performed, and the recoveries are 97. 0% and 110%, respectively.
Keywords:Carboxyethyl germanium sesquioxide  Electrothermally atomised atomic absorption spectrometry  Silicon  
本文献已被 CNKI 等数据库收录!
点击此处可从《高等学校化学研究》浏览原始摘要信息