首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Chromatographic impulse response technique with curve fitting to measure binary diffusion coefficients and retention factors using polymer-coated capillary columns
Authors:Kong Chang Yi  Funazukuri Toshitaka  Kagei Seiichiro
Institution:Faculty of Environment and Information Sciences, Yokohama National University, 79-7 Tokiwadai, Hodogaya-ku, Yokohama 240-8501, Japan. kong@ynu.ac.jp
Abstract:The theoretical basis of a Gaussian-like approximate solution was applied to a chromatographic impulse response technique with curve fitting for measuring binary diffusion coefficients and retention factors using a polymer-coated capillary column. The formulae were derived for evaluating both the accuracy of the approximate solution and the sensitivity of the parameters. The validity of the solution also was confirmed experimentally for pulse injection of phenol in acetone into supercritical carbon dioxide flowing at 313.15 K and 11.6-28.6 MPa. Potential sources for experimental errors of the method are discussed.
Keywords:Impulse response technique  Curve fitting  Binary diffusion coefficient  Retention factor  Mathematical modeling  Transient response
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号