首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Design and performance of a desktop time-of-flight mass spectrometer for analyzing metal ions
Authors:Guo Changjuan  Gao Wei  Huang Zhengzu  Zhu Hui  Fu Zhong  Dong Junguo  Cheng Ping  Zhou Zhen
Institution:School of Chemistry and Environment, South China Normal University, Guangzhou 510006, China.
Abstract:We have described a home-made desktop orthogonal-injection time-of-flight (O-ToF) mass spectrometer combining a collisional cooling system. This O-ToF consists of a simple electrospray ion source, an atmosphere-vacuum interface, an area of transmission, including a radio-frequency only quadrupole (RF- only quadrupole, RFQ) as a collisional cooling cell and an orthogonal ToF mass analyzer. In order to detect ions of small m/z value, such as small metal ions, the RFQ has been improved to weaken the mass discrimination effect against low mass ions. Metal salt solutions were used in the experiment. The system has shown a satisfactory resolving power in the spectra (m/Δm = 3500), a good mass stability, a limit of detection of 80 fg and a mass accuracy of 48 ppm. The dynamic range is found to be from 10(-8) mol L(-1) to 10(-5) mol L(-1), allowing the semi-quantitative analysis of metal ions.
Keywords:
本文献已被 PubMed 等数据库收录!