首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Parts per trillion sensitivity for ethane in air with an optical parametric oscillator cavity leak-out spectrometer
Authors:von Basum Golo  Halmer Daniel  Hering Peter  Mürtz Manfred  Schiller Stephan  Müller Frank  Popp Alexander  Kühnemann Frank
Institution:Institut für Lasermedizin, Universit?t Düsseldorf, Universit?tsstrasse 1, D-40225 Düsseldorf, Germany.
Abstract:Spectroscopic detection of ethane in the 3-microm wavelength region was performed by means of a cw optical parametric oscillator and cavity leak-out. We achieved a minimum detectable absorption coefficient of 1.6 x 10(-10) cm 1/square root of Hz, corresponding to an ethane detection limit of 6 parts per trillion/square root of Hz. For 3-min integration time the detection limit was 0.5 parts per trillion. The levels are to our knowledge the best demonstrated so far. These frequency-tuning capabilities facilitated multigas analysis with simultaneous monitoring of ethane, methane, and water vapor in human breath.