  • Abstract:A dielectric barrier discharge (DBD) was used as a new atmospheric optical emission detector for the determination of trace nitrogen in pure argon gas in this work. The whole system was composed of an ac ozone generation device for power supply, a six-way valve, a laboratory-built DBD device and a USB2000 charge coupled device (CCD). Trace nitrogen in argon was detected at nitrogen molecular emission line of 337 nm. This method features with several advantages: atmospheric working condition, low power consumption (≤ 12 W), simple and cheap instrumentation, fast response and high sensitivity and accuracy. Under the optimized conditions, the limits of detection can be down to 34 ppb.
    Keywords:Dielectric barrier discharge   Nitrogen   Argon   Molecular emission spectrometry   Portable instrumentation