直热式毛细管柱对直馏汽油组成的分析
引用本文:刘晓微,杨海鹰,陆婉珍.直热式毛细管柱对直馏汽油组成的分析[J].色谱,1998,16(3):198-202.
作者姓名:刘晓微  杨海鹰  陆婉珍
作者单位:中国石化总公司石油化工科学研究院
摘    要:采用柱内电阻丝加热升温的直热式13X分子筛薄层毛细管填充柱分析汽油的族组成。分析时间比炉温加热短50min,定量分析结果与炉温加热和多孔层开管柱基本相同。保留时间的相对偏差小于3%,直馏汽油中绝大部分组分定量结果的相对偏差小于5%。

关 键 词:毛细管气相色谱法  气相色谱柱  汽油分析  

Analysis of Hydrocarbon Composition in Gasoline with Direct Heating Packed Capillary Column
Liu Xiaowei,Yang Haiying and Lu Wanzhen.Analysis of Hydrocarbon Composition in Gasoline with Direct Heating Packed Capillary Column[J].Chinese Journal of Chromatography,1998,16(3):198-202.
Authors:Liu Xiaowei  Yang Haiying and Lu Wanzhen
Abstract:A packed glass capillary column with an electric resistance wire inside is a new type of column. It can be directly used for temperature programming without GC oven. Satisfactory results have been obtained with this new column type for gas analysis. In this paper, straight run gasoline is analyzed by using this column type. Stationary phase is made by coating 13X molecular sieve on 102 Packing. Put the stationary phase into a glass tubing with an electric resistance wire inside and draw them into packed capillary column with a metal core. Two metal unions are specially designed on each end of the column. The union links up the glass capillary column and a quartz capillary. Electric power is applied through outside of the metal union which is in close contact with the inside heating wire. Graphite ferrules are used for their endurance of high temperature. The temperature of the column is programmed by gradual increase of electric voltage applied. Baseline compensation is used in analyzing straight run gasoline by the direct heating column. The apparent temperature can be observed through the thermocouple bound to the column. The apparent and actual temperature can be correlated by the retention time of components. The actual temperature of direct heating column is usually above the apparent value. The repeatability of retention time for most component of six runs is less than 3%. The relative deviation of quantitative results for each component in straight run gasoline is less than 5%. The analysis time is shorter by direct heating mode than that by oven heating. Increasing the coat thickness of 13X molecular sieve will increase the column efficiency, but the elution temperature is increased when the thickness of molecular sieve coat increased. Thirty percent 13X molecular sieve loaded on 102 is used for the above analysis.
Keywords:capillary gas chromatography  gas chromatographic column  hydrocarbon analysis