首页 | 本学科首页   官方微博 | 高级检索  
     


Flow injection analysis coupled with HPLC and CE for monitoring chemical production processes
Authors:N. Dantan  W. Frenzel  S. Küppers
Affiliation:1. Fachgebiet Luftreinhaltung KF2, TU-Berlin, Institut für Technischen Umweltschutz, Stra?e des 17 Juni 135, 10623, Berlin
2. Chemical Development, Schering AG, Müllerstr. 170-178, 13342, Berlin
Abstract:Summary In chemical and pharmaceutical production plants process control is often performed by plant personal near the process (at-line). Alternatively, spectroscopic procedures like near infrared may be coupled to the process using in-line or on-line interfaces. When the analytical problem cannot be solved by these established approaches chromatographic techniques can be directly coupled to the production process. An application of chemical reaction monitoring which is difficult to perform is reported. The analytical problems are solved by HPLC and CE. Both techniques in the application presented need an on-line derivatization step prior to the analysis. Flow Injection Analysis (FIA) was applied for this sample preparation step. For the online coupling to the commercially available CE-instrument a special sample vial was developed. The application shows sufficient reproducibility and analysis speed for near real-time monitoring. of the process. Dedicated to Professor Dr. Heinz Engelhardt on the occasion of his 65th birthday.
Keywords:Column liquid chromatography  Capillary electrophoresis  Flow injection analysis  Hyphenated methods  Process monitoring
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏