首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Photoelectrochemical immunosensor for label-free detection and quantification of anti-cholera toxin antibody
Authors:Haddour Naoufel  Chauvin Jérôme  Gondran Chantal  Cosnier Serge
Institution:Laboratoire d'Electrochimie Organique et de Photochimie Rédox (CNRS UMR 5630), Institut de Chimie Moléculaire de Grenoble FR CNRS 2607, Université Joseph Fourier, BP 53, 38041 Grenoble Cedex 9, France.
Abstract:We demonstrate herein a newly developed photoelectrochemical immunosensor for the determination of anti-cholera toxin antibody by using a photosensitive biotinylated polypyrrole film. The latter was generated by electro-oxidation of a biotinylated tris(bipyridyl) ruthenium(II) complex bearing pyrrole groups. The photoexcitation of this modified electrode potentiostated at 0.5 V vs SCE, in the presence of an oxidative quencher, pentaaminechloro cobalt(III) chloride (15 mM), led to a cathodic photocurrent. As a result of the affinity interactions, a layer of biotinylated cholera toxin was firmly bound to the functionalized polypyrrole film via avidin bridges. The resulting modified electrodes were tested as immunosensors for the detection of the corresponding antibody from 0 to 200 microg mL(-)(1). The antibody concentration was measured through the decrease in photocurrent intensity resulting from its specific binding onto the polymeric coating, the detection limit being 0.5 microg mL(-)(1).
Keywords:
本文献已被 PubMed 等数据库收录!