首页 | 本学科首页   官方微博 | 高级检索  
     


Online Monitoring of Aluminium in Drinking Water with Fluorimetric Detection
Authors:A. Alonso-Mateos  M. J. Almendral-Parra  Y. Curto-Serrano  F. J. Rodríguez-Martín
Affiliation:(1) Departamento de Química Analítica, Nutrición y Bromatología, Faculty of Chemistry, University of Salamanca, Plaza de la Merced s/n, 37008 Salamanca, Spain
Abstract:We report a procedure for the online monitoring of aluminium in drinking water by flow injection analysis. The reaction used is the formation of a complex with morin. Under the working conditions, this can be accomplished in an ethanol-rich hydroalcoholic medium, which modifies the fluorescent characteristics of the complex, allowing the determination of aluminium concentrations higher than 3.1 μgl−1, with a linear application range between 2 and 250 μgl−1, an R.S.D. of 2.3% (n = 10, 120 μgl−1) and a sampling frequency of 90 h−1. The method can thus be considered one of the most sensitive and fastest for the continuous determination of aluminium. In the presence of anionic surfactants, the sensitivity of the determination is increased. In this form, aluminium is detected at concentrations higher than 2.8 μgl−1, with a linear application range of 2–50 μgl−1. The procedure was applied to the analysis of aluminium in drinking, river, and underground water. Under the proposed working conditions, only Fe(III), fluoride and phosphates interfere. The interference of Fe(III) can be avoided with hydroxylamine and that of phosphates and polyphosphates by acid digestion of the samples.
Keywords:Aluminium  Morin  Flow injection analysis  Fluorimetry  Drinking water
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号