A miniaturized FIA system for the determination of residual chlorine in environmental water samples.
Authors:Jiye Jin  Yasutada Suzuki  Naoyoshi Ishikawa  Toyohide Takeuchi
Institution:Department of Chemistry, Faculty of Science, Shinshu University, 3-1-1 Asahi, Matsumoto, Nagano 390-8621, Japan. jin@gipac.shinshu-u.ac.jp
Abstract:The present study was aimed at developing a portable miniaturized FIA system with a two-electrode based amperometric detector for on-site monitoring of residual chlorine in environmental samples. The electrode potential control and current data collection were conducted with a home-designed potentiostat with a size of 150 x 100 x 40 mm and the weight of 350 g. It was confirmed that the amperometric response increased in proportion to the chlorine concentration up to 5 microg/ml, and the detection limit for chlorine was 0.05 microg/ml. Besides a drastic reduction in the size of the analytical system, such miniaturization achieved the advantages of increased speed, minimal mobile-phase consumption, and high sensitivity. The proposed method was successfully applied to the determination of residual chlorine in samples of pool water and tap water. The results were well agreed with those obtained by the N,N-diethyl-p-phenylenediamine (DPD) photometric method.