Kinetic titration with differential thermometric determination of the end-point
Authors:Sajó I
Institution:Iron and Steel Research Institute Budapest, Hungary.
Abstract:A method has been described for the determination of concentrations below 10(-4)M by applying catalytic reactions and using thermometric end-point determination. A reference solution, identical with the sample solution except for catalyst, is titrated with catalyst solution until the rates of reaction become the same, as shown by a null deflection on a galvanometer connected via bridge circuits to two opposed thermistors placed in the solutions.