首页 | 本学科首页   官方微博 | 高级检索  
     


Direct monitoring of drug degradation by easy ambient sonic-spray ionization mass spectrometry: the case of enalapril
Authors:Amaral Phellipe H  Fernandes Raquel  Eberlin Marcos N  Höehr Nelci F
Affiliation:Faculty of Medical Sciences, Department of Clinical Pathology, University of Campinas- UNICAMP, Campinas, SP, Brazil.
Abstract:Using enalapril maleate as a test case, the ability of ambient mass spectrometry, namely, via easy ambient sonic‐spray ionization mass spectrometry (EASI‐MS), to perform direct monitoring of drug degradation has been tested. Two manufacturing processes were investigated (direct compression and wet granulation), and the formation of degradation products was measured via both EASI‐MS and high‐performance liquid chromatography with ultraviolet detection for a total period of 18 months. Both techniques provide comparable results, which indicate that direct analysis by ambient mass spectrometric techniques presents a viable alternative for drug degradation monitoring with superior simplicity, throughput, and reliability (no sample manipulation), and comparable quantitative results. In terms of qualitative monitoring, the full mass spectra with intact species provided by EASI‐MS allow for comprehensive monitoring of known and unknown (or unexpected) degradation products. Copyright © 2011 John Wiley & Sons, Ltd.
Keywords:ambient mass spectrometry  sonic‐spray ionization  drugs  drug degradation products
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号