首页 | 本学科首页   官方微博 | 高级检索  
     


Separation of Photosynthetic Pigments by High-performance Liquid Chromatography: Comparison of Column Performance,Mobile Phase,and Temperature
Affiliation:Ma Chung Research Center for Photosynthetic Pigments, Universitas Ma Chung, Malang 65151, East Java, Indonesia
Abstract:High-performance liquid chromatography (HPLC) has been commonly used as method of separating and identifying photosynthetic pigments such as chlorophylls and carotenoids because of such advantages as speed, high resolution and sensitivity. In this technique, high separation relies largely on the type of column material. This study compared the efficiency of five reverse-phase columns, C8, C18, C18 monolithic, π-NAP, and cholester, for separation of photosynthetic pigments at several fixed conditions of mobile phase and temperature. This investigation also analysed the parameters of ΔtR and tR ratio for selected pigments and resolution for structural isomers, such as α- and β-carotene. Among above columns tested, cholester column is suitable for separation of pigments not only for a broad range of polarity, but also for hydrophobic pigments in a simple mobile phase. This finding can help in the selection of column and HPLC parameters in separating photosynthetic pigments.
Keywords:Cholesteryl bonded  HPLC column  monolithic packing  particulate packing  photosynthetic pigments  reverse phase.
本文献已被 ScienceDirect 等数据库收录!