Metrologically based procedures for the temperature, heat and heat flow rate calibration of DSC
Authors:S M Sarge  W Hemminger  E Gmelin  G W H Höhne  H K Cammenga  W Eysel
Institution:(1) Physikalisch-Technische Bundesanstalt, Bundesallec 100, D-38116 Braunschweig, Germany
Abstract:Metrologically based measuring procedures and evaluation methods are recommended as guidance for practical temperature, heat and heat flow rate calibration of DSC instruments which are largely independent of instrumental, test and sample parameters. The relevant terms are defined, the measuring procedures and evaluation methods described, calibration materials and their characteristic data stated and guidance for the sample handling provided. Reference is made to three extended papers on calibration. The recommendations were developed by the working group lsquoCalibration of Scanning Calorimetersrsquo of the German Society of Thermal Analysis (GEFTA).Recommendation of the working group lsquoCalibration of scanning calorimetersrsquo of the Gesellschaft für Thermische Analyse e.V. (GEFTA)
Keywords:calibration material  calibration procedure  DSC  heat calibration  heat flow rate calibration  standards  temperature calibration
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏