首页 | 本学科首页   官方微博 | 高级检索  
     


Study of the responses of a gas chromatography—reduction gas detector system to gaseous hydrocarbons under different conditions
Authors:Xu-Liang Cao   C. Nicholas Hewitt  K. S. Waterhouse
Affiliation:

Institute of Environmental and Biological Sciences, Lancaster University, Lancaster LA1 4YQ, UK

Abstract:The response of a reduction gas detector (RGD) to C2-C6 alkenes, C2-C6 alkanes, isoprene and benzene has been investigated using gas chromatography (GC) with a packed column. The RGD is considerably more sensitive to alkenes than is the flame ionization detector. The detection limit of this present GC/RGD system for alkenes is about 0.01 ng. It has much greater sensitivity to alkenes than to alkanes. Its sensitivity increases with increasing HgO bed temperature, but its selectivity towards alkenes decreases at the same time. The selectivity of the RGD may not be significant for much heavier molecules. The sensitivity of the RGD is inversely proportional to the carrier gas flow rate through the HgO bed. The baseline of the system increases significantly with increasing oven temperature.
Keywords:Gas chromatography   Reduction gas detector
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号