Liquid chromatography tandem mass spectrometry method for determination of bisoprolol in human plasma using d5‐bisoprolol as the internal standard
Authors:Gang‐yi Liu  Wei Wang  Jing‐ying Jia  Chuan Lu  Yan‐mei Liu  Meng‐qi Zhang  Yun Liu  Shui‐jun Li  Chen Yu
Affiliation:1. Central Laboratory, Shanghai Xuhui Central Hospital,966 Huaihai Middle Road, Shanghai 200031, China;2. Department of Emergency ICU, Shanghai Xuhui Central Hospital,966 Huaihai Middle Road, Shanghai 200031, China
Abstract:A simple, reliable and sensitive liquid chromatography tandem mass spectrometry (LC‐MS/MS) protocol was developed and validated for quantification of bisoprolol in human plasma. The sample was pretreated with a simple procedure of protein precipitation and an isotope‐labeled d5‐bisoprolol was used as internal standard. The chromatographic separation was performed on a Capcell Pak C18 MG III column (100 mm × 2.0 mm, 5 µm). The protonated ion of the analyte was detected in positive ionization by multiple reaction monitoring mode. The mass transition pairs of m/z 326.3 → 116.3 and m/z 331.3 → 121.3 were used to detect bisoprolol and the internal standard, respectively. Linearity, accuracy, precision, recovery, matrix effect, dilution test and stability were evaluated during method validation over the range of 0.5–100 ng/mL. The validated method was successfully applied to analyze human plasma samples in a bisoprolol bioavailability study. Copyright © 2009 John Wiley & Sons, Ltd.
Keywords:bisoprolol  liquid chromatography tandem mass spectrometry  human plasma